Medical/Commercial Transient Generators

Meet conducted immunity requirements in IEC/EN standards such as IEC 60601 and EN 550022. The EMC Shop is an ISO 17025 accredited company. Contact us to rent of buy Medical/Commercial Transient Generators, most equipment is in stock and can ship same day.

Medical/Commercial Transient Generators There is 1 product.

Subcategories

  • Combination Wave/Surge Generators

    Combination Wave/Surge generators are required to meet IEC/EN 61000-4-5ANSI (IEEE) C62.41 and other conducted immunity standards requiring a surge EMC test. The EMC Shop stocks all types of Combination Wave/Surge generators for compliance and pre-compliance testing. Rent or buy Combination/Surge generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.



    Overview Combination/Wave Surge Generators

    Combination Wave/Surge generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.

     

    Common Combination Wave/Surge Tests

      • IEC/EN 61000-4-5 specifies two Combination Wave pulses: Open Circuit Voltage Short Circuit Current
      • A common combination wave test is the 1.2/50 & 8/20
      • Another common wave test is the 10/700 & 5/320, or otherwise known as telecom wave
      • The four standard test levels , excluded of EUT: 0.5, 1, 2, 4 kV (0.25, 0.5, 1, 2 kA) Surge Test Environment

    Combination Wave/Surge Generator Videos





    Get Application Support, Pricing, Lead Time and More: Contact Form

  • Electrical Fast Transient EFT/Burst Generators

     

    EMC/EMI Testing Services              Automotive EMI/EMC Testing              OTA/Antenna Pattern Measurement

     

    Electrical Fast Transient (EFT)/Burst generators are required for conducted immunity testing standards, such as IEC/EN 61000-4-4.  Electrical Fast Transient EFT/Burst events are small electrical surges caused by a local source, typically an electrical switch, it can occur within a product itself as well as from the data or power lines in the facility.



    The EMC Shop stocks most all types of EFT/Burst equipment for compliance and pre-compliance testing. Rent or buy stand alone Electrical Fast Transient (EFT)/Burst generators or try a multifunction generator to add SurgePower Fail and other electrical disturbance tests.

    Get Application Support, Pricing, Lead Time and More: Contact Form



    Electrical Fast Transient EFT/Burst Generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.



    Common Electrical Fast Transient EFT/Burst Tests

    A typical burst has a single pulse rise time/duration of 5 ns/50 ns from a source impedance of 50 Ω

    Bursts of 15 ms duration with a repetition rate of 5 kHz (or 100 kHz) are applied every 300 ms

    The standard defines four basic test levels, all with repetition frequencies 5 or 100 kHz

    The level voltage depends on equipment being tested:

    Power ports: 0.5 , 1, 2, 4 kV

    Signal and control ports: 0.25, 0.5, 1, 2 kV

  • Surge & Burst/EFT Generators

    Conducted immunity testing is a major requirement to attain CE marking for products. Multiple types of electrical disturbances are required to be tested and certified to pass the internal compliance standards, EFT/Burst standard EN/IEC 61000-4-4 and Surge (combo wave) standard IEC/EN 61000-4-5 are available in most standard EMI test sets made today. The EMC Shop stocks all types of Surge & Burst/EFT Generators for compliance and pre-compliance testing. Rent or buySurge & Burst/EFT Generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: Contact Form



    Transient generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.

    Common Transient Tests

    Combination Wave/Surge:

    • IEC/EN 61000-4-5 specifies two Combination Wave pulses: Open Circuit Voltage Short Circuit Current
    • A common combination wave test is the 1.2/50 & 8/20
    • Another common wave test is the 10/700 & 5/320, or otherwise known as telecom wave
    • The four standard test levels , excluded of EUT: 0.5, 1, 2, 4 kV (0.25, 0.5, 1, 2 kA)

    Burst/EFT:

    • A typical burst has a single pulse rise time/duration of 5 ns/50 ns from a source impedance of 50 Ω
    • Bursts of 15 ms duration with a repetition rate of 5 kHz (or 100 kHz) are applied every 300 ms
    • The standard defines four basic test levels, all with repetition frequencies 5 or 100 kHz
    • The level voltage depends on equipment being tested:
    • Power ports: 0.5 , 1, 2, 4 kV
    • Signal and control ports: 0.25, 0.5, 1, 2 kV

  • Surge, Burst/EFT, Ring Wave, Dips & Interrupts

    Accomplish all common conducted immunity testing to conform with EMC (electromagnetic compliance) standards for CE marking in one multi-purpose instrument. Generate electrical disturbances for Surge (Combination wave pulse 1.2/50 - 8/20 μs), Burst/EFT (5/50 ns), and Ringwave (0.5 μs/100 kHz) as well as pass power quality compliance standards by simulating AC dips, drops, and interruptions. The EMC Shop stocks all types of multifunction generators for compliance and pre-compliance testing. Rent or buy generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: CONTACT FORM



    Multifunction generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.



    Common Combination Wave/Surge Tests

    • IEC/EN 61000-4-5 specifies two Combination Wave pulses: Open Circuit Voltage Short Circuit Current
    • A common combination wave test is the 1.2/50 & 8/20
    • Another common wave test is the 10/700 & 5/320, or otherwise known as telecom wave
    • The four standard test levels , excluded of EUT: 0.5, 1, 2, 4 kV (0.25, 0.5, 1, 2 kA)

    Common (EFT)/Burst Generator Tests

    • A typical burst has a single pulse rise time/duration of 5 ns/50 ns from a source impedance of 50 Ω
    • Bursts of 15 ms duration with a repetition rate of 5 kHz (or 100 kHz) are applied every 300 ms
    • The standard defines four basic test levels, all with repetition frequencies 5 or 100 kHz
    • The level voltage depends on equipment being tested:
    • Power ports: 0.5 , 1, 2, 4 kV
    • Signal and control ports: 0.25, 0.5, 1, 2 kV

    Ring Wave Immunity Generator Tests

    For compliance, the ring wave has a measured 0.5 us rise time and 100 kHz oscillation frequency This impulse is commonly used for applications related to power distribution and railway environments.

  • Magnetic Field Generators

    Conducted immunity testing is a major requirement to attain CE marking for products. Magnetic pulse field strengths conforming to EN/IEC 61000-4-8 and EN/IEC 61000-4-9 are the most common EMI magnetic test. The EMC Shop stocks magnetic field generation equipment for compliance and pre-compliance testing. Rent or buy magnetic field generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: CONTACT FORM

    Common Magnetic Field Generation Tests

    • magnetic disturbances at power frequencies 50 Hz and 60 Hz related to
    • residential and commercial locations
    • industrial installations and power plants
    • medium voltage and high voltage sub-stations

     

  • Flicker & Harmonics Generators

    Flicker and harmonics generators are required to perform compliance testing to IEC 61000-3-2, 3-3, 3-11, and 3-12. The EMC Shop has power sources available for rent or sell to perform this testing. The EMC Shop stocks all types of Flicker and harmonics generators for compliance and pre-compliance testing. Rent or buy Flicker and harmonics generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: CONTACT FORM



    Flicker and harmonics generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.



    Common Flicker and Harmonics Generator Tests

     

  • 3-Phase CDN's

    For large commercial products requiring three phase power at 240V, 480V or 600V. An additional CDN will required with the transient generators to complete IEC 61000-4-4, IEC 61000-4-5ANSI C62.45 and other conducted immunity testing. The EMC Shop stocks 3 Phase coupling/decoupling networks for rental or purchase.

    Get Application Support, Pricing, Lead Time and More: Contact Form


    Three Phase CDN Diagram per IEC 61000-4-5

    Coupling/Decoupling Network Diagram for 3-Phase EUT per IEC 61000-4-5

    Three Phase CDN Coupler Applications

    EN/IEC 61000-4-5

    EN/IEC 61000-4-4

    ANSI (IEEE) C62.45

    EN/IEC 61000-4-11

    EN/IEC 61000-4-29

     

  • Data Lines Surge CDN's

    The EMC Shop is an ISO 17025 accredited company. Contact us to rent of buy Data Lines Surge CDN's .

    Get Application Support, Pricing, Lead Time and More: Contact Form



    Data Lines Surge CDN's Standards:

    EN/IEC 61000-4-5

    EN/IEC 61000-4-4

    ANSI (IEEE) C62.45

    EN/IEC 61000-4-11

    EN/IEC 61000-4-29

     

Showing 1 - 1 of 1 item
    • Multiple configurations in stock
    • Rentals ship same day
    • ISO 17025 calibration available
    • 3 Phase Coupling and Data Line Clamps Available
    Request Quote/More Info
Showing 1 - 1 of 1 item