Description
The Teseq 9206 CFL has been designed to interface between a transducer (antenna, line network) and measuring equipment (receiver, spectrum analyzer). The CFL 9206A assists in preventing damage to input circuits of measuring equipment from high level transients often encountered during conducted emission testing. The CFL 9206A incorporates a band pass filter and 10 dB high power attenuator. The maximum input power ratings are for signals within the CFL 9206A passband.
Features & specifications Teseq CFL 9206:
- Insertion loss: 10 dB ±0.5 dB
- Frequency response: 9 kHz to 30 MHz
- Maximum continuous input: 36 dBm at 25°C (for signal within the passband)
- Input temperature de-rating linear: 30 dBm at 50°C
- Maximum DC voltage on: 400 Volts
- Input (1 second): 42 dBm
- Peak power (non-repetitive <10 μs): 4 kW
- Maximum output power (for +42 dBm input): 15 dBm
- Maximum input level (no compression): 15 dBm
- 50 Hz rejection: 100 dB
- System impedance: 50 Ω
- Connectors: BNC
- Operating temperature: 0 to 50°C
- Size (W x D x H): 140 mm x 70 mm x 25 mm
- Weight: approx. 185 g
Videos
Custom Field
Weight approx. 185 g
Dimensions (W x D x H): 140 mm x 70 mm x 25 mm
(5.6 x 2.8 x 1) in.
Frequency Range 9 kHz to 30 MHz
Input 36 dBm at 25deg. C (for signal within the passband)
Input temperature de-rating linear: 30 dBm at 50deg. C
Insertion loss 10 dB ±0.5 dB
Max input power 4 kW
Maximum Continuous Power 15 dBm
RF Connector BNC
Voltage Rating 400 Volts