Amplifier Research FP6001 E-Field Probe


No reviews yet Write a Review
Application/Test Type:
Radiated Immunity
Field Strength Range:
0.5 to 800 V/m
Frequency Range:
100 kHz to 6 GHz
Standards Met:
IEC 61000-4-3
Interfaces:
Fiber Optic
Battery Life:
10 Hours
Manufacturer:
ETS-Lindgren

 

SKU:
sku-1294-1

 


Getting Started...

Check out our Google Reviews

A.R. FP6001 Overview

The AR Field Probe is the same as the ETS-Lindgren HI-6005

[split]

  • ON (I) - When the power switch is in the I position, an internal 4.8 VDC Nickel-Metal Hydride (NiMH) battery provides power to the probe, and the power indicator LED blinks.

  • OFF (0) - When the power switch is in the O position, the probe is inactive. When not in use, turn the probe off to prolong battery life.
ETS-Lindgren HI-6005 Electric Field Probe

 

ETS-Lindgren HI-6005 Electric Field Probe Specifications
Frequency Range 100 kHz to 6 GHz
Frequency Response 500 kHz to 2 GHz +1.0, -2.5 dB
2 to 5.5 GHz +3.5, -4.0 dB
5.5 to 6 GHz +2.0, -6.0 dB
Frequency Response with Correction 100 kHz to 6 GHz ± 0.9 dB
Dynamic Range 0.5 - 800 V/m (Single Range)
Resolution 0.01 V/m
Isotropic Deviation ± 0.5 dB @ 400 MHz
Linearity ± 0.5 dB @ 27 MHz

Standard Accessories:

  • FP6001
  • Hard case
  • Certificate of Calibration
  • HI-4413USB
  • 4.8 NiMH Charger

A.R. FP6001 Key Considerations

Manufacturer:
ETS-Lindgren
Application/Test Type:
Radiated Immunity
Field Strength Range:
0.5 to 800 V/m
Frequency Range:
100 kHz to 6 GHz
Standards Met:
IEC 61000-4-3
Interfaces:
Fiber Optic
Battery Life:
10 Hours